Technical Specifications
| Parameter | Specification |
|---|---|
| Probe Model | 330101-00-66-10-02-CN |
| Probe Tip Material | Polyphenylene Sulfide (PPS) |
| Probe Case Material | AISI 303 or 304 Stainless Steel |
| Cable Type | 75Ω triaxial, FEP insulated (standard) or PFA (ETR) |
| Connector Type | Gold-plated ClickLoc connector, tool-free installation |
| Cable Tensile Strength | 330 N at probe end; 270 N at cable junction |
| Recommended Mount Torque | Forward: 11.2 N·m; Reverse: 7.5 N·m |
| Total Length | 1.0 meter |
| Weight | 0.323 kg |
| Probe Temperature Range | Standard: -52°C to +177°C; ETR: -52°C to +218°C |
| Cable Temperature Range | Standard: -52°C to +177°C; ETR: -52°C to +260°C |
| Proximitor Sensor Temp Range | Operating: -52°C to +100°C; Storage: -52°C to +105°C |
| Recommended Gap Voltage | -9 Vdc (~1.27 mm gap) |
| Linear Range | 2 mm (80 mils), from 0.25 mm to 2.3 mm |
| Incremental Scale Factor (ISF) | 7.87 V/mm (200 mV/mil) ± 5% (1m and 5m systems) |
| Power Requirements | -17.5 Vdc to -26 Vdc, max 12 mA current |
| Output Resistance | 50 Ω |
| Min Cable Bend Radius | 25.4 mm |
| Recommended Shaft Diameter | Minimum: 50.8 mm; Preferred: ≥76.2 mm |
| Frequency Response | 0 to 10 kHz, +0, -3 dB |
| EMI/RFI Immunity | CE compliant, no metal housing required |
FAQ
FAQ
Q1: Is this probe system fully interchangeable?
A1: Yes, all components (probe, extension cable, Proximitor sensor) are fully interchangeable without the need for matched calibration.
Q2: Can it be used with legacy 3300 5 mm or 8 mm series components?
A2: Yes, but the system’s performance will default to that of the non-XL components.
Q3: What is the maximum linear range of the system?
A3: The linear range is 2 mm (80 mils), spanning from 0.25 mm to 2.3 mm from the target.
Q4: Are any special tools needed for installation?
A4: No special tools are required. ClickLoc connectors are designed for finger-tight installation.
Q5: Is this product suitable for high-temperature applications?
A5: Yes, the ETR version supports temperatures up to +218°C for the probe and +260°C for the cable.
Q6: Can the Proximitor sensor be DIN-rail mounted?
A6: Yes, it supports both DIN-rail and traditional panel mounting.
Q7: Does the probe system provide protection against electromagnetic interference (EMI/RFI)?
A7: Yes. The 3300 XL Proximitor Sensor offers excellent EMI and RFI immunity, allowing installation in fiberglass housings without additional shielding, and complies with CE standards.
Q8: Does this product comply with international certifications?
A8: Yes, it meets EMC, RoHS, ATEX, IECEx, CSA, and other major international standards.